Test Method
For flash point determination of biodiesel, distillate fuels, new lubricating oils, residual fuel oils, cutback residua, used lubricating oils, mixtures of petroleum liquids with solids, petroleum liquids that tend to form a surface film during testing.
Specifications
Conforms to the specifications of:
ASTM D93 Procedure A, B and C; IP 34; ISO 2719; DIN EN 22719; JIS K2265; NF M 07-019
Detection System: Thermocouple and Ionization Ring
Temperature Range: Ambient - 405°C
Heating Rate: In accordance to ASTM D93 procedure A, B and C
Stirring Rate: 0 to 300 ± 5 RPM
Cycle Time: 5 minutes as per rapid fan cooling system Ignition Test Frequency: User selectable on per method basis
Cooling:
Integrated Dual Fan System: First (1) directly to cup, Second(2) to cool environment around test cup
Barometric Pressure: Automatic Barometric Pressure Correction
Methods & Data Storage: Limited only by PC hard disk drive space
Temperature Calibration: Three Types come standard:
2 Point Sample Temperature
Multipoint (>2) Temperature Calibration
CRM Calibration – Built-In Software function to automatically correct as per CRM tested
Communication: LIMS Connectivity via Ethernet and RS 232 ports. Two (2) USB ports (1 – Frontm 2 – Back)
Security: Three (3) level password protection
Safety: Fire prevention, detection, and suppression systems standard