Specification
Accuracy and repeatability equivalent to ASTM approved CFR engines test
methods (ASTM D2699, D2700)
Sample Holder: Sealed, cubical glass container (75mm optical path length)
Sample Volume: 8 Ounces (approx. 225 mL)
Operating Temperature Range: 7°C - 38°C - Pre-calibrated for unleaded & ethanol-blended gasoline (Analyzer can be calibrated for up to 8 additional fuel types.) Battery operated (6 AA batteries)
Description
Measures octane number via near-infrared (NIR) transmission spectroscopy utilizing 14 near-infrared emitting diodes with narrow bandpass filters, a silicon detector system, and a fully integrated microprocessor. Simple octane number determination requires three easy steps: sampling a background signal, acquiring two absorption spectra of the gas sample, and then acquiring a second background signal. Analyzer is pre-calibrated for unleaded gasoline and ethanol-blended fuels, and can be calibrated for up to eight additional fuel types. The analyzer is small, lightweight, and operates on "AA" batteries or AC. Before each reading, the unit standardizes itself to assure accuracy. The octane number is printed with time and date on the built-in printer. All data can be downloaded via the RS232 port to an external computer.
Test Method
Determines the Pump Octane Number (AKI), Research Octane Number (RON), and Motor Octane Number (MON) of unleaded gasoline, ethanol blended gasoline, leaded gasoline and Cetane Number for diesel fuels.
Included Accessories
Features
Additional Accessories
Electrical Requirements
115V 60Hz
220-240V 50/60Hz
Alat Laboratorium Umum
Untuk keterangan lebih lanjut, silahkan download PDF ini :