Spesifikasi Real-Time Spectrum Analyzer Rigol RSA3015N - 1.5 GHz with Vector Network Analysis
Rigol RSA3015N - 1.5 GHz Real-Time Spectrum Analyzer with Vector Network Analysis
1.5 GHz Real-Time Spectrum Analyzer w/10MHz RTBW - VNA Capable
Real time spectrum analyzer with vector network analysis
Frequency range: 9kHz to 1.5 GHz
VNA frequency range: 100kHz to 1.5 GHz
Includes tracking generator
Ultra-Real technology
Multiple trigger modes and trigger masks
Density, spectrogram and other display modes
PC software options
10.1” capacitive multi-touch screen with gesture supports
USB, LAN, HDMI and other communication and display interfaces
OCXO option is no longer sold separately. Please see OCXO version of this model for factory installed clock
RSA3015N
Frequency Range 9 kHz to 1.5 GHz
Internal Reference Frequency
Reference Frequency 10 MHz
Accuracy ±[(time since last calibration × aging rate) + temperature stability + calibration accuracy]
Initial Calibration Accuracy Standard: <1 ppm
Option OCXO-C08: <0.1 ppm
Temperature Stability 0℃ to 50℃ , with the reference 25℃
Standard: <0.5 ppm
Option OCXO-C08: <0.005 ppm
Aging Rate Standard: <1 ppm/year
Option OCXO-C08: <0.03 ppm/year
SSB Phase Noise 20℃ to 30℃,fc = 500 MHz
Carrier Offset 1 kHz: <-90 dBc/Hz (typical)
10 kHz: <-100 dBc/Hz, <-102 dBc/Hz (typical)
100 kHz: <-100 dBc/Hz, <-102 dBc/Hz (typical)
1 MHz: <-110 dBc/Hz, <-112 dBc/Hz (typical)
Displayed Average Noise Level (DANL) attenuation = 0 dB, sample detector, trace averages ≥ 50, tracking generator off, normalized to 1 Hz, 20℃ to 30℃, input impedance = 50 Ω.
Preamp off 9 kHz to 100 kHz: <-120 dBm (typical)
100 kHz to 20 MHz: <-135 dBm, <-140 dBm (typical)
20 MHz to 2.7 GHz: <-138 dBm, <-141 dBm (typical)
2.7 GHz to 3.0 GHz: <-136 dBm, <-141 dBm (typical)
3.0 GHz to 4.5 GHz: <-136 dBm, <-140 dBm (typical)
Preamp on 100 kHz to 20 MHz: <-152 dBm, <-160 dBm (typical)
20 MHz to 2.7 GHz: <-158 dBm, <-161 dBm (typical)
2.7 GHz to 3.0 GHz: <-156 dBm, <-161 dBm (typical)
3.0 GHz to 4.5 GHz: <-154 dBm, <-159 dBm (typical)
Level Measurement Uncertainty 95% confidence level, S/N > 20 dB, RBW = VBW = 1 kHz, preamp off, attenuation = 10 dB, -50 dBm < input level ≤ 0 dBm, fc > 10 MHz, 20℃ to 30℃
Level Measurement Uncertainty <1.0 dB (nominal)
Network Analyzer