phone Phone
+6281290773137
close
menu

CV. Andalan Prima Sejahtera

sql
GENRAD VI-700 PRECISION VOLTAGE & CURRENT SOURCE
Hover to Zoom

Jual GENRAD VI-700 PRECISION VOLTAGE & CURRENT SOURCE



Spesifikasi GENRAD VI-700 PRECISION VOLTAGE & CURRENT SOURCE

VI-700 PRECISION VOLTAGE & CURRENT SOURCE

Features

  • Voltage output range of 100 µV to 20 V or 200 V
  • Current output range of 0.1 µA to 20 mA or 200 mA
  • 3½ digit resolution, 0.75 LSD accuracy
  • Manual thumbwheel operation or programmable digital input, IEEE-488 or RS232 interface
  • May be used as a component in a closed loop control system

Various features and options make the VI-700 an effective choice for many manual and programmable voltage and current source applications.

The VI-700 Voltage and Current Source is a versatile precision dialable or remotely programmable voltage and current source. Its three ranges provide up to 20 V full scale (200 V with the HV option), and down to 100 mV resolution; and 20 mA full scale down to 0.1mA resolution. Both voltage and current outputs are available simultaneously. The output is set by a range selector and thumbwheel switches in the LOCAL mode, or by a digital input at the rear panel. The REMOTE mode control can be a RS232 or IEEE-488 interface.

This voltage and current source offers many full performance features; fulfilling a variety of laboratory needs. Its wide outputrange (up to 200 V), very low output impedance (5 mΩ, typical), low noise and ripple, and 200 mA load current capability suit the VI-700 to a variety of applications, such as simulation, calibration, A/D andD/A converter evaluations, and product development.  Front panel overload indicator lamps warn of exceeding voltage load or current compliance limits. This feature assures the user that the output is within specifications without having to make any measurement or computations. 

Specification

Voltage
Output Ranges200 mV, 2 V, 20 V;
(200 V with HV option).
Resolution100 µV, 1 mV, 10 mV;
(100 mV with HV option).
Accuracy±(75 ppm + 50 µV); 200 mV range;
±(75 ppm + 0.25 mV), 2 V range;
±(75 ppm + 2.5 mV), 20 V range;
at 23°C, no load, after warmup.
Output Current  70 mA, (200 mA with HP option) maximum
load up to 20 V; (10 mA for 200 V range with
HV option); with LED over load in di ca tor.
Output Impedance5 m Ω typical, 40 m Ω for 20 V range.
Noise and Ripple50 µVrms on 200 mV and 2 V ranges;
50 µVrms on 200 mV and 2 V ranges;
100 µVrms on 20 V range.
Tempco±10 ppm/°C typical;
±15 ppm/°C max.

 

Current
Output Ranges200µA, 2 mA, 20 mA,
200 mA (with HI option).
Resolution0.1µA, 1µA, 10µA;
100µA (with HI option).
Accuracy±(75 ppm + .05µA); 200µA;
±(75 ppm + 0.25µA), 2 mA;
±(75 ppm + 2.5µA), 20 mA; 23°C, no load.
Compliance Voltage0-20 V; with LED overload indicator.
Noise8-20 nA.

 

General
Digital Input31/2 digit parallel BCD or IEEE-488
in terface; a front panel push-button
selects REMOTE or LOCAL op er a tion.
IsolationFloating outputs, optically isolated from
digital inputs; voltage and current out-
puts may be used in either polarity.
Power105-125 V or 210-250 V;
Requirements50-60 Hz; 5 W.
Dimensions21.6 cm W x 12.6 cm H x 22.9 cm D
(8.5 x 2.8" x 9.0" ) for Standard Model; "
21.6 cm W x 8.7 cm Hx 23.5 cm (8.5 "
x 3.44 x 9.25" ) for IEEE-488 interface option. "
Weight2.3 kg (5 lb).