Horiba OCMA-500 Oil Content Analyzer
OCMA-500 to measure concentration of oil contained in drainage and environmental water.
After injecting the water sample and solvent, all you have to do is press the start button,and the system will automatically conduct the monitoring operation from oil extraction to sample measurement and draining. With no more troublesome operations like opening/closing the drainage valve, monitoring is speeded up.In addition, the color graphic LCD and the backlit extraction tank have improved operability.
Specification
- Model: OCMA-500
- Product name: Oil content analyzer
- Measurement method: Solvent extraction - non-dispersive infrared absorption analysis method
- Measured objects: Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm
- Measurement range: 0 mg/L to 200 mg/L
- Resolution For mg/L: 0 to 99.9: 0.1, 100 to 200: 1
- For mg/g, mg/kg, mg/PC: 0 to 9.99: 0.1, 10.0 to 99.9: 0.1, 100 to 200: 1
- Repeatability: 0 mg/L to 9.9 mg/L: ±0.3 mg/L.10.0 mg/L to 99.9 mg/L: ±2.1 mg/L.100 mg/L to 200 mg/L: ±5 mg/L. * For standard liquids
- Display method: 3.5 inches, 320×240 dots, Backlight Color graphic LCD
- Calibration method: Zero, span calibration
- Amount of test sample required: Sample water : Solvent = 2:1
- Extraction solvent: S-316
- Amount of extraction solvent required: 8 mL
- Extraction method: Built-in extractor
- Ambient operating temperature: 0°C to 40°C (no condensation)
- Power supply: AC 100 V to 240 V ±10%, 50/60 Hz
- Power consumption: AC 100 V: Approx. 60 VA, AC 240 V: Approx. 90 VA
- External dimensions: 342 (H) × 200 (W) × 313 (D) mm
- Mass: Approx. 7 kg
- External output: Output to a USB memory stick
Functions
- 300-item data memory
- Self error determination
- Stabilized measurement value display
- Clock
- Compliant with ASTM D7066-4
Oil Analyzer