sql

Horiba Oil Content Analyzer Model OCMA-550

Rp. 100
Update Terakhir
04 Okt 2024
Negara Asal
Japan
Pembelian Min
1 Unit
Kategori
HORIBA

Horiba Oil Content Analyzer ModelOCMA-550

OCMA-550 to measure residual oil on components and concentrations of oil adhered on solids such as soil.

Measurement can be easily made only by injecting the extracted water sample into the attached cell and setting it to the equipment.

This model features a simple design which allows opening/closing of the door to setting of cell with just one hand.

This is best for measurement of extracted samples such as evaluation of residual oil on components and measurement of oil contained in food.

Specification

  • Model: OCMA-550
  • Product name: Oil content analyzer
  • Measurement method: Solvent extraction - non-dispersive infrared absorption analysis method
  • Measured objects: Substances extracted from sample water into solvent and having infrared absorption near a wavelength from 3.4 μm to 3.5 μm
  • Measurement range: 0 mg/L to 200 mg/L
  • Resolution: For mg/L 0 to 99.9: 0.1, 100 to 200: 1For mg/g, mg/kg, mg/PC 0 to 9.99: 0.01, 10.0 to 99.9: 0.1, 100 to 200: 1Abs.: 0 to 1.000 Abs.: 0.001 Abs.
  • Repeatability: 0 mg/L to 9.9 mg/L: ±0.4 mg/L ±1 dig.10.0 mg/L to 99.9 mg/L: ±2.0 mg/L ±1 dig.100 mg/L to 200 mg/L: ±4 mg/L ±1 dig. * For standard liquids
  • Display method: 3.5 inches, 320×240 dots,Backlight Color graphic LCD
  • Calibration method: Select each optionally zero calibration and span calibration
  • Extraction solvent: S-316
  • Amount of extraction solvent required: Approx. 6.5 mL
  • Extraction method: Using the extraction solvent, and extracted manually outside the product
  • Ambient operating temperature: 0°C to 40°C (no condensation)
  • Power supply: 100 V to 240 V AC ±10%, 50/60 Hz
  • Power consumption: 100 V to 240 V AC, Approx. 60 VA
  • External dimensions: 195 (H) × 253 (W) × 293 (D) mm
  • Mass: Approx. 5 kg
  • External output: Output to a USB memory stick

Functions

  • 300-item data memory (measurement history)
  • Stabilized measurement value display
  • Self error determination
  • Clock
  • Conversion of measurement units
  • Compliant with ASTM D7066-4

Oil Analyzer

Lihat Selengkapnya

Untuk keterangan lebih lanjut, silahkan download PDF ini :

Ulasan

/5
Ulasan
5
4
3
2
1

Foto dari Pembeli

Ulasan dari Pembeli

Filter
Semua
Dengan Foto
5
4
3
2
1
Minta Penawaran
Produk Siap di Pesan
Rp 0
1