Specificatioon


Measured Parameters and Basic measurement accuracy *: Any two of 14 parameters measured and displayed simultaneously, user selectable


ParameterMeasurement RangeFastMediumSlow
Cs, Cp00000 .01 fF to 9.999999 F±0.5%±0.25%±0.05%
Ls, Lp0000.001 nH to 99.99999 H±0.5%±0.25%±0.05%
D.0000001 to 99 .99999±0.005±0.0025±0.0005
Q.0000001 to 999999.9±0.005±0.0025±0.0005
Z. Rs, Rp, ESR,Xs000.0001 mΩ to 99 .99999 MΩ±0.5%±0.25%±0.05%
Y, Gp, Bp00000 .01 µS to 9.999999 MS±0.5%±0.25%±0.05%
Phase Angle-180.0000° to +179.9999°±1.8°±0.9°±0.18°

*At optimum test signal levels, frequencies, DUT values and without calibration uncertainty. Capacitance (Cs/Cp), Inductance (Ls/Lp), Resistance (Rs/Rp), Dissipation (D) and Quality (Q) Factors, Impedance Z , Admittance Y , Phase Angle (0-), Equivalent Series Resistance (ESR), Conductance (Gp), Reactance (Xs), Susceptance (Bp) Note: s = series, p = parallel, ESR equivalent to Rs


Test Frequency: Range: 10 Hz to 2 MHz, continuous Resolution:6 digits or 0.05 Hz


Frequency Accuracy: ±(0.01% + 0.10 Hz)


Measurement Speed: Fast: 120 meas/sec, Medium: 16 meas/sec - 8 meas/sec below 150kHz, Slow: 2 meas/sec - 1 meas/sec below 150kHz


Ranging: Automatic, Range Hold or user selectable


Trigger: Internal (automatic), External (LAN, RS232, GPIB, USB) and Manual Start Button


AC Test Signal:


Voltage: 20 mV to 5.0 V (open circuit) up to 500kHz in 5 mV steps 20 mV to 1.0 V (open circuit) 500kHz-1MHz in 5 mV steps 20 mV to 0.5 V (open circuit) >1MHz in 5 mV steps


Current: 250 μA to 100 mA (short circuit) in 50 μA steps Max. Compliance 3V < 500kHz.


Source Impedance: 25Ω, 400Ω, 6.4kΩ, or 100kΩ, range dependent


DC Bias Voltage: Internal: 2.0 V External: 0 to +/-200V


Sweep Result: Primary parameter vs. frequency, voltage or current Graphical or Tabular Format Up to 250 measurement points per sweep


Sequencing Result: Displays up to 6 sequential test results, primary and/or secondary


AutoAcc: Automatic calculation and display of overall instrument accuracy for selected settings, test conditions, and device under test


Interfaces: USB Host Port, USB Type B, GPIB, LAN, RS-232, Handler


Charged Capacitor Protection: √8/C for Vmax < 250 V; √2/C for Vmax < 1000V C = Capacitance in farads of the device under test


Measurement Delay: Programmable from 0 - 1000 ms in 1 ms steps


Averaging: Programmable from 1 - 512 Median value mode


Data Storage: USB Host Port USB1.1 Compliant, ASCII format


Program Storage: Memory internal, USB Host Port, ASCII format


Mechanical: Bench mount with tilt bail Dimensions: (w x h x d): 16 x 6 x 14in (410 x 150 x 360mm) Weight: 17 lbs (8kg) net, 23 lbs (10.5kg) shipping


Safety: IEC61010-1: 2010 CAT II, Pollution Degree 2


EMC: 2014/30/EU