Test Method
For determination of evaporation loss of lubrication grease at temperatures from 93-316° C. A controlled flow of heated air is passed over the sample for a specified period. Evaporation loss is measured by the change in sample weight during the test.
Specifications
Conforms to the specifications of:
ASTM D2595, D2878**with accessory oil sample cup installed
Capacity:2 samples
Temperature Range:200 to 600°F (93 to 316°C)
Sample Temperature Control:Type: microprocessor digital control
Exit Air Temperature Control:Two 0-500W variable control heaters and type K thermocouples (order K29320/K29329 Digital Thermometer separately)
Air Flow Control:Two externally mounted flowmeters maintaining 2L/min flowat standard temperature and pressure